Place a
blob of gel either on end of probe or on tummy, place
probe gently on tummy, move probe to find your baby’s
heart beat (once detected, visual display will flash FHR
(usually between 120-160bpm). Enjoy listening to your
baby’s heartbeat, be assured that his/her heartbeat is
normal.
It is perfectly safe to use this product as many times
and as long as you like each time. However we do
recommend that you limit each use to 10 minutes so as
not to interrupt baby’s normal routine in the womb.
IMPORTANT: This product is not
meant to substitute your regular gynae/ob visits. Please
check with your gynae/ob IMMEDIATELY if the Doppler
displays an abnormal FHR in RED or if you are unable to
detect baby’s heartbeat.
About the product - Technical Specifications.
This product has received USA FDA510-k approval
Safety: Complies with: IEC 60601-1:1988+A1:1991+A2:1995
According to manufacturer - CE technically approved
■ Anti-electroshock Type: Internally powered equipment.
■ Anti-electroshock Degree: Type B applied part
■ Harmful Liquid Proof Degree:
■ Main unit Ordinary equipment (sealed equipment without
liquid proof).
■ Probe: Prevent from water splashing, degree of
protection: IPX4.
■ Degree of Safety in Presence of Flammable Gases:
Equipment not suitable for use in
■ presence of flammable gases.
■ Working System: Continuous running equipment.
■ EMC: Group I Class B.
■ Suitable Using Range: Suitable for use after the 12th
week of pregnancy.
Physical Characteristic
■ Size: 135mm (Length) ×92mm (Width) ×29 (Height) mm
■ Weight: About 245g (including batteries)
Environment Working:
■ Temperature: +5℃~+40℃
■ Humidity: ≤80%
■ Atmospheric Pressure: 70 kPa~l06kPa Transport and
Storage:
■ Temperature: -10℃~+55℃
■ Humidity: ≤93%
■ Atmospheric Pressure: 50 kPa~l06kPa Display: 45mm×25mm
LCD display
FHR Performance
■ FHR Measuring Range: 50~240BPM (BPM: beat per minute)
■ Resolution: 1BPM
■ Accuracy: ±2BPM
■ Power consumption:<0.8W
■ Auto Shut-OFF: After 1 minute no signal, power off
automatically.
■ Battery Type Recommended: Two pieces of 1.5 V DC
battery (SIZE AA LR6).
Probe:
■ Nominal Frequency: 2.0MHz
■ Working Frequency: 2.0MHz±10%
■ P-: <0.5MPa
■ Iob: <10 mW/cm2
■ Ispta: <50mW/cm2
■ Ultrasonic Output Intensity: Isata<5 mW/cm2
■ Ultrasonic Output Power: P<10 mW
■ Working Mode: Continuous wave Doppler
■ Effective Radiating Area of Transducer: 157mm2±15%
|